Version 5.2

Features, changes and bug fixes introduced in version 5.1.0.79.

These features and fixes build on the previous 5.1 releases.

Minor Releases

Click here to see the release notes for minor releases since this initial 5.0 release.

Features

  • added progress bar for task list loading process

  • currently executed and scheduled tasks are highlighted with a colored background

  • improved logging

enviro TOC

  • FLEXDILUTION:

    • There is a warning if the dilution factor is out of range (minimum=1, maximum=250)

    • It is not possible to deactivate the feature while a task is running

  • CLD:

    • in case that the CLD is not running, it is possible to activate it by clicking on the CLD status message in the system status window

    • in the device settings there are two buttons to turn the CLD on and off

    • if the inlet valve (gas) is deactivated the CLD will be turned off

  • For standards there is a tolerance value available, samples will not run if the standard is not within tolerance.

  • enviroTOC control methods are hidden by default.

  • Initialize and Wake Up methods are also working if there is an error shown - in some cases an initialization of the system can solve the shown error.

  • Add to batch button has been removed for enviroTOC.

  • BCB firmware upgrade for multivalve movement optimisation.

  • If device is sleeping, only "Wake Up" method can be started by clicking the "Wake Up" button in enviroTOC tab.

  • There is a warning message if the calibration method doesn´t match the chosen task method.

  • The maintenance manager can be opened by clicking on the system status attribute to reset counters easily, in the maintenance manager window there are visual progress bars to see the current value of the consumables.

  • With a service login calibrations can be exported and imported.

 

IRMS

  • There were added N2 methods because of new EAS CARBSORBER:

    • EA Analysis N2

    • EA Analysis N2 (iso TOC)

    • EA Analysis N2 Flush (iso TOC)

  • GHG ribbon tab added

Bugfixes

  • Connection failure between lyticOS and GC 8890 has been fixed.

  • enviroTOC:

    • Editing a calibration in an opened task list was not possible in some cases.

    • In solid mode it was not possible to choose position 1 right after position 60 in task list.

    • NPOC method did not use the acidification volume which was set in the method parameters.

    • Suitability test method names were missing in some cases.

    • TIC Solid missing parameter bug has been fixed.

  • Generated report of GC results did not contain all graphs.

  • Methods which belong to not connected systems are not available and will not be displayed anymore.

  • Minor bugfixes