Minor Releases
4.0.0.7546 (Initial 4.0 Release)
Click here to see the release notes for the initial 4.0 release.
4.0.1.7575
- Improvements to the isoprime visION / precisION vacuum initialization logic to ensure that the system is correctly initialized after a firmware upgrade.
- Correctly report the gauges model and serial number information in isoprime visION / precisION readbacks.
- Support sending of customized error reports from within Log Messages window and Connection Failed dialog
- Fix right click - View Info on an acquired species dataset (acquisition or scan) to display used tuning parameters
- Fix NICM digital inputs
- Add global setting to allow methods to run if required devices used by a method are not detected. Users will see a gray warning icon instead of the usual orange warning icon.
- Add method block to check if a device is configured for a system
- Add support for dynamic properties within the NICM device settings for use within custom methods
- Improvements to dynamic properties data type editor
4.0.2.7588
- Data Grid Field Chooser layout has been updated to allows groups of fields to all be selected or unselected
- Fix intermittent issues with the Method / Dashboard designer when dragging connected blocks outside of the currently displayed view
- IonVantage importer now allows multiple folders to be selected at the same time.
- IonVantage importer will look at the method file, parameter file and Map file to try and detect the species type.
- Within a GC or EA batch, the adding or clearing of exclusion windows can be applied to all selected samples.
- Within an EA batch, ensure that exclusion windows are correctly applied to subsequent species (after the first species).
4.0.3.7614
- Peak Maps can now be stored in lyticOS global settings making them easily exportable or importable into batches
- Peak Maps editor improvements (Add, Grouping, Rename Species)
- Global settings manager improved to open settings automatically in relevant dialogs
- GC5 Cal Controllers now reports a duty cycle readback
- Allow the iso FLOW acid pump maximum strokes to be configurable within the device settings
- Dilution types can now be grouped
- German language pack updates
- Method log now records summary information (e.g. Stable, Not Stable, etc.)
- Dashboards now load in a correct size which take into account invisible items
- Manage Methods dialog no longer disables the connected system window
- Report BCB firmware version information with the Status panel
- New firmware to ensure that the filament does not automatically turn on again after the vacuum interlock is restored
- Improvements to the firmware update process
- Improvements to source status messages and behavior when the source vacuum or cover interlocks are broken and restored
4.0.4.7645
- More reliably detect if the system is idle (for automation)
- Only display the Low Diluter Pressure warning if the inlet motor is in position 1 (EA)
- Stop running orphaned blocks within parallel method workflows
- Added method block to get a device name for a particular device type
- Further improvements to firmware update process
- Extend the BCB Comms utility to optionally report firmware log information in a new tab
- Speed up dashboard loading time and performance
- Add units to the GC Method Editor (XML View)
- Ensure that all results are available for use by Calculated columns within GC Batch Processing
- Ensure that calculated column results are included in batch reports
- Fix delayed tasks not working
4.0.5.7706
- Improvement to when the "Reinitialise Source" status message is displayed
- Manage Methods / Flows dialog now automatically scrolls to show the selected Method / Flow
- Ensure that all dashboard items "visibility" property is ignored in design mode (always shown in design mode)
- Improve method flow editor performance for large flows
- Improve schematic flow path emulation. Multi-port components now display emulated flow path colors.
- Dashboard items can now bind to system or device specific dynamic settings (also editable by methods). Effectively allowing control of dashboard items from methods.
- Panel visibility can now be changed from within a method
- Improvements to German language pack. Ensure that more text is translatable.
- Improvements to translation tools. Monitor most accessed text to prioritize translation. Be able to open the Language Update Utility easily from any window. Be able to change the current user language from any window. Language packs can now be enabled / disabled.
- Quick access to common functions now available from the "lyticOS" icon in the top left hand corner of any window (settings, utilities, help / about, log out).
- BUG FIXED: Delayed tasks not running. Sometimes after delaying a task and then starting the task list, if you make any other changes to the task list, the delayed task would not run.
4.0.6.7728
- Prevent incorrectly configured email settings from stopping the task list when trying to send a failed task warning notification
- Fix the custom rack creation wizard for the Gilson auto sampler
- Improvements to the method flow connection path text positioning
- Performance improvements
- The BCB Send Command method blocks are now available to customer licenses
4.0.0.7546 (Initial 4.0 Release)
Click here to see the release notes for the initial 4.0 release.